User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: Honeywell UDC 5000 4 - 20mA DATE: 22-Nov-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 13 NUMBER OF LINES: 108 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- R P F W 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITION ÍÍ 1.004 DISP Press UUT LOWR DISP button. 1.005 DISP Press SET UP until DISPLAY function is reached. 1.006 DISP Press FUNC to display the DECIMAL configuration. 1.007 DISP Use up and down arrow keys to set decimal reading 1.007 DISP to XX.XX format. 1.008 DISP Press UUT LOWR DISP button. 1.009 DISP Connect a Honeywell 62.5ê resistor accross input 1.009 DISP terminals 22 and 23. 1.010 DISP ÿ Connect the 5700A and the UUT as follows: 1.010 DISP ÿ [27][91]1m 5700A TO UDC 5000 1.010 DISP ÿ OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ INPUT 22 1.010 DISP ÿ OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ INPUT 23 1.011 HEAD {} 1.012 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.013 HEAD {º FULL SCALE AND LINEARITY TEST º} 1.014 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.015 HEAD {}ÍÍ FULL SCALE AND LINEARITY TEST ÍÍ 1.016 JMP 2.001 1.017 EVAL 2.001 5700 4mA S 2W 2.002 MEMI Use keyboard to enter actual UUT reading in Percent. 2.003 ACC 0.0pctrh 0.0003U 2.004 MEME 2.005 MEMC pctrh 0.05U 4mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 3.001 5700 5.6mA S 2W 3.002 MEMI Use keyboard to enter actual UUT reading in Percent. 3.003 ACC 10.0pctrh 0.0003U 3.004 MEME 3.005 MEMC pctrh 0.05U 5.6mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 4.001 5700 7.2mA S 2W 4.002 MEMI Use keyboard to enter actual UUT reading in Percent. 4.003 ACC 20.0pctrh 0.0003U 4.004 MEME 4.005 MEMC pctrh 0.05U 7.2mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 5.001 5700 8.8mA S 2W 5.002 MEMI Use keyboard to enter actual UUT reading in Percent. 5.003 ACC 30.0pctrh 0.0003U 5.004 MEME 5.005 MEMC pctrh 0.05U 8.8mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 6.001 5700 10.4mA S 2W 6.002 MEMI Use keyboard to enter actual UUT reading in Percent. 6.003 ACC 40.0pctrh 0.0003U 6.004 MEME 6.005 MEMC pctrh 0.05U 10.4mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 7.001 5700 12mA S 2W 7.002 MEMI Use keyboard to enter actual UUT reading in Percent. 7.003 ACC 50.0pctrh 0.0003U 7.004 MEME 7.005 MEMC pctrh 0.05U 12mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 8.001 5700 13.6mA S 2W 8.002 MEMI Use keyboard to enter actual UUT reading in Percent. 8.003 ACC 60.0pctrh 0.0003U 8.004 MEME 8.005 MEMC pctrh 0.05U 13.6mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 9.001 5700 15.2mA S 2W 9.002 MEMI Use keyboard to enter actual UUT reading in Percent. 9.003 ACC 70.0pctrh 0.0003U 9.004 MEME 9.005 MEMC pctrh 0.05U 15.2mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 10.001 5700 16.8mA S 2W 10.002 MEMI Use keyboard to enter actual UUT reading in Percent. 10.003 ACC 80.0pctrh 0.0003U 10.004 MEME 10.005 MEMC pctrh 0.05U 16.8mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 11.001 5700 18.4mA S 2W 11.002 MEMI Use keyboard to enter actual UUT reading in Percent. 11.003 ACC 90.0pctrh 0.0003U 11.004 MEME 11.005 MEMC pctrh 0.05U 18.4mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 12.001 5700 20mA S 2W 12.002 MEMI Use keyboard to enter actual UUT reading in Percent. 12.003 ACC 100.0pctrh 0.0003U 12.004 MEME 12.005 MEMC pctrh 0.05U 20mA #! Test Tol 0.1, Sys Tol 0.0003, TUR 333.333 (>= 4.00). 13.001 HEAD {} 13.002 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.